General Information
CFULL automatic endoscope washer-disinfector is researched and developed according to standard ISO 15883-4. It adopts automatic micro processor controlling.The product is made of washing tank and pipeline system and can be used with many kinds of disinfectant such as: glutaraldehyde, peracetic acid, OPA, chlorinedioxide and electrolyzed oxidizing water. The treatment for endoscope includes leaking test, blocking test, wash, enzyme wash, rinse, disinfection, final rinse,alcohol drying and air drying.
Features
Leaking Test Function
The leaking test will finish before the endoscope contacts the liquid and will also keep testing during the operation. The system gives out visual and audible signals and will terminate operation as soon as the leaking valueexceeds the setting value.
Blocking Test Function (Optional)
The blocking test will start before the endoscope contacts the liquid and after the operation. The system gives out visual and audible signals and will
terminate operation as soon as the pipe is blocked.
100% Wash-Disinfection
Omnibearing wash-disinfection
The tank equipped with rotating nozzle and spray arm to wash the outer surface of the endoscope, in the meanwhile, the circulating water flow keeps flushing the endoscope from inside
Advantages
Endoscope Pressure Booster Pump
It is used for washing of tiny pipe, such as air/water injection nozzle, to reduce the formation probability of bacteria biofilm.
Build-In Water Treatment Function
It is equipped with build-in double-level (0.45 um and 0.2 m) water filter which could filter out the water impurity to guarantee the disinfection result
Sterile Water Rinse
To avoid secondary contamination caused by unsanitary water, the endoscope will be rinsed by sterile water which filtered by 0.2 um filter after disinfection.
Drying Function
It is used for drying the inner and outer surface of endoscope. Cfull equipped with air drying and alcohol drying for user to order
Self-disinfection Function
The self-disinfection program can be started during maintenance or after the termination of operation. The function is designed to disinfect for wash
disinfection tank (includes 0.2 um filter) and pipeline.
Most Speedy Wash And Disinfection
An endoscope can be wash-disinfected within 15mins by Cfull which dramatically improves the endoscope turnover efficiency.
Note: Specific wash-disinfection time depends on the type of disinfectant.>Perfect medical workers protection
Auto Door With Foot Switch
The automatic visual glass door is adopted for observing the operation inside the tank. The door can be opened by gently kicking the foot switch.
Totally Enclosed
Cfull wash-disinfection tank adopts totally-enclosed structure to avoid sanitizer odor leakage so that protect the medical workers.