Flectrical Requirements 220VAC/110VAC,50 or 60Hz,200VA
Operation Interface Available in Engish,French,Spanish,German and Chinese
Volume 0.1 ml to syringe capacity in 0.1 ml increments
Pressure Preset limit:300 psi default,user settable from 50 to 350 psi in 1 psi increments
Display:Real-time pressure curve
Change threshold:Stop injection if pressure changing suddenly
Alarm:Voice alarm and stop injection when over preset limit
njection or Scan Delay 0 to 3599 seconds in 1 sec.increments
Autofill Settable flow rate from 3-8ml/sec.,volume from 10ml to syringe capability
Multi-Phase 1-8 phases per injection
Pause Phase 0~999 sec.in 1 sec.increments
Hold Capability More than 30 minutes
Storage Capabilities 1000 protocols
Expel air locked Unable to inject unless expelling air
Test injection Settable flow rate from 0.1-5ml/sec.,volume from 0.1-10ml
Heat maintainer 35~38°C(optional)
Hand switch Hand switch control pause and continue injection (optional)
Emergency stop function Clicking anywhere on the touch screen of the console or pressing any button on the injector arm to stop protocol injection
Dimension Injector Arm and Pedestal:L:22.8”(58.0cm),W:12.5°(20.1cm),H:50.2“(127.5cm)
Console:L:4.55°(11.55cm),W:12.68"(31.8cm),H:9.77°(24.81cm)
Main controller:L:11.9°(30.2cm)W:7.4°(9.1cm)H:4.7°(12cm)
Weight Injector Arm and Pedestal:16.5Kg
Console:4.1Kg
Main Controller:6.2Kg