Disposable Invasive Blood Pressure Transducer IBP DPT Transducer for Abbott JIBPT-01-YP-S
$5.80 / Parcel Min.order:1000

About this item:

Disposable Pressure Transducer is used in ICU Anesthesiology and enable to consist and accurate reading of arterial and venous blood pressure measurement. Available with flushing of 3cc/hr, 30cc/hr or without flush valves. - Infusion set - Pressure monitoring line - Stopcock etc - Infusion set - Pressure monitoring line - Stopcock etc - Medical disposable

Support Payment Term:

HSBC Hong Kong

paypal

Alibaba Pay

Western Union

USD

EUR

GBP

SGD

HKD

CNH

CAD

MXN

BRL

JPY

THB

MOP

AUD

NZD

PLN

CZK

HUF

RON

CHF

SEK

NOK

DKK

TRY

AED

SAR

ILS

ZAR

$0.00

Quantity

Discover similar items

Page 1 of 2

Model NO.:JIBPT-01-YP

Logo Printing:with Logo Printing

Material:High Polymer Material and Electronic Chips

Quality Guarantee Period:Three Years

Group:Adult/Chidren

Name:Disposable Pressure Transducer Kit

Properties:Medical Materials & Accessories

Usage:Hospital, Clinic, ICU Room

Certificate:CE/ISO13485/TUV

Feature:with Flush Device

Packing:100PCS/Carton/53*46*45cm

Model Number:Single /Double/Triple Channel Kit

Product Features:Compatiable Adapter

Function:to Measure The Pressure of Blood

Stock:Yes

Transport Package:Sterile Blister Box,

Specification:Compatible with Abbott Single Channel

Trademark:JCR

Origin:Shenzhen, China

HS Code:9018909991

Production Capacity:4000000/Month

Product Description

Properties

Medical Materials & Accessories

Product Name

Abbott disposable IBP transducer

Brand Name

JCR

Model Number

JIBPT-01-YP


Main technical parameters and working environment requirements:

Disposable Pressure Transducer

Barometric pressure

70~106Kpa

Relative humidity

10~90% (non-condensing)

Operating pressure range

-50 ~ + 300 mmHg

Sensitivity

5.0μV/V/mmHg ±3%

Nonlinearity and hysteresis

full-scale reading±1.5%

Input impedance

1200Ω ~3200Ω

Zero pressure offset

-20mmHg~+20mmHg

Thermal offset shift

≤±0.3mmHg/°C

Offset drift

After warm-up for 20 seconds, drift within 2 mmHg within 8 hours

Thermal span shift

≤±0.1%/°C

Frequency response

Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz

Defibrillator withstand

Defibrillation depends on the final connection of the equipment.

Leakage current

Leakage current is dependent on the final connection of the equipment

Overpressure load

-400~+4000mmHg

Shock resistance

Withstand falling from one meter for three times

 

Light sensitivity

When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.

Storage and transport environment requirements

- 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.

Contact with the human body

≤168h

Intra-system velocity

Flow

2-5mL/h

Intra-system refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.

Product shelf-life

3 years