Manufacturing Compatible Nellcor Oximax Disposable Sensor Medical SpO2 Probe
$1.40 / Parcel Min.order:100
Application method This product shall be used in combination with the required model: For the matching compatible equipment, the degree of protection against electric shock of its applied part shall be at least F-type applied part Operation principle, function and structure The patient sensor end of the product is attached on the patient's index finger (Thumb, middle finger or others, great toes), and the device connector is connected to compatible monitors or oximeter with SpO2 extension cord.

Support Payment Term:

HSBC Hong Kong

paypal

Alibaba Pay

Western Union

USD

EUR

GBP

SGD

HKD

CNH

CAD

MXN

BRL

JPY

THB

MOP

AUD

NZD

PLN

CZK

HUF

RON

CHF

SEK

NOK

DKK

TRY

AED

SAR

ILS

ZAR

$0.00

Quantity

Discover similar items

Page 1 of 2

Application method
This product shall be used in combination with the required model: For the matching compatible equipment, the degree of protection against electric shock of its applied part shall be at least F-type applied part

Operation principle, function and structure

The patient sensor end of the product is attached on the patient's index finger (Thumb, middle finger or others, great toes), and the device connector is connected to compatible monitors or oximeter with SpO2 extension cord. The product adopts 660nm red light and 905nm Infrared light as incident light source, tests light intensity through tissue to obtain SpO2 value as well as pulse rate.

Basic Info.

Model NO.:J92S003P

Ethylene Oxide Sterilization:Ethylene Oxide Sterilization

Quality Guarantee Period:Two Years

Group:Adult

Logo Printing:With Logo Printing

Usage:for Patient Monitor

Tpye:Disposable

Weight:0.026kg

Function:Blood Oxygen Saturation

Properties:Medical Materials & Accessories

Colour:Blue

Transport Package:1PCS/Bag

Specification:900mm

Trademark:JCR

Origin:Guangdong, China

HS Code:854890009

Production Capacity:200000PCS/Month

Indications
The product is indicated for single use to collect and transmit signal of SpO2 and pulse rate of patient, with compatible monitors or oximeter. The product is only used by experienced medical professional or under their directions.

Specifications

Measurement Wavelengths and Output Power

Red: 660±5nm @ 2.0 mw maximum average

 Infrared: 905nm±5nm @2.0 mw maximum average Oxygen Saturation Declared;

Accuracy Range

±3 % of 70%~79% SpO2; ±2% of 80%~89% SpO2; ±1% of 90%~100% SpO2;

Pulse Rate Range

30~245 BPM;

Accuracy

 ±3 bpm of 150 bpm ~245 bpm; ±2 bpm of 60 bpm~149 bpm; ±1 bpm of 30 bpm~59 bpm;

Response time

 ≤30s