DX - 8000 - a chassis suspension parameters suspension support structure l

electric multi-function photography system stand (100 cm, 180 cm)/decubitus conversion l

projection position: Stand 1000 mm / 1800 mm, recumbent position and oblique l

X bulb lateral movement range: mobile travel l

X 1800 mm + 50 mm ball tube longitudinal motion range: mobile travel l

X 1800 mm + 50 mm ball tube vertical movement range: Mobile travel l

X 1500 mm + 50 mm ball tube rotation range: rotation Angle 0 ° plus or minus 2 ° ~ + 90 ° plus or minus 2 ° l does the sternum craft bracket vertically range: 500 mm to 1700 mm + / - 5% l

chest radiograph craft bracket rotation Angle range: 0 ° plus or minus 2 ° ~ 90 ° + 2 ° l

made a film from the SID: 1000 mm to 1800 mm

lie a film from SID: 1000 mm l l l

bed structure mobile photography

bed

bed surface: long x 2200 mm x 800 mm wide l

height: 670 mm

material: carbon cellulosic polymer high voltage generator

High voltage generator

Power: 50 kw

KV photography: 40 KV ~ 150 KV

Photography mA: 10 ~ 640 mA

MAs: 0.1 mAs ~ 500 mAs

Inverter frequency: ≥40kHz

ARP: support

Control mode: console operation

X-ray tube assembly

Focus size: 0.6/1.2

Focus power: 20kW/50kW-50HZ

Anode heat capacity: 300kHU

Heat capacity of tube sleeve: 1250kHU

Anode target Angle: 12°

Maximum tube voltage: 150kV

Direct Digital Imaging Detector (import)

Material: amorphous silicon cesium iodide implant

Size: 14 x 17 inches

Pixels: 7 megapixels

A/D: 16 bits

DQE: 70%

Pixel matrix: 2448×2984

Pixel size: 143 m

Imaging time: ≤3s

Ultimate resolution: ≥3.7LP/MM

Data transmission mode: mobile wire

Limit beam splitter

Specification: AC24V / 100 w

Control mode: manual

Minimum exposure field: <50mm×50mm

Radiation field (SID=1000mm) :430mm×430mm

Average exposure brightness: >140Lux;

Illumination: >4:1

Intrinsic filtration: 1.0mmAL equivalent

Lead equivalent of shield blade: 3mMPB (intermediate blade: 2mMPB)

The filter wire grid

Grid density: ≥80L/cm

Grid ratio: ≥ 10:1

Grid focal length: 100cm/180cm

The workstation

CPU: Kujong I5 quad-core 3.2ghz

Memory: 4 gb

Hard disk capacity: 1TB

DVD drive: DVD drive

System interface: USB, RS232, DVI/VGA/HDMI/DP, COM serial port, PCI/PCI-E

Monitor: 22 "LCD

Operating system: Windows 7 Pro legal edition

Workstation software functions  

User operation: patient information registration, multi-condition inquiry of location patient information, patient appointment, emergency treatment,

Image processing functions: window width/window, local automatic window width/window bits, preset width/window, plus or minus as inversion, image rollovers, rotating, image magnification and roaming edge enhancement, image interpolation, local amplification, restore the original image, text/digital annotation, image scale line measurement, Angle measurement, brightness/contrast adjustment

Image printing: DICOM printing, paper printing, image contrast/brightness adjustment, multi-image piecing printing, film size selection, print queue control.