Original Beckman Refurbished Medical Brand Au480 Auto Chemistry Analyzer
$13382.35 / Parcel$16058.82 Min.order:1
DDP Area
Put the power of versatility to work for your lab with a smart and efficient clinical chemistry analyzer solution. The AU480 chemistry analyzer is the ideal primary clinical chemistry analyzer for low- to mid-volume hospitals and laboratories, or dedicated specialty chemistry or STAT analyzer for larger laboratories.

Support Payment Term:

HSBC Hong Kong

paypal

Alibaba Pay

Western Union

USD

EUR

GBP

SGD

HKD

CNH

CAD

MXN

BRL

JPY

THB

MOP

AUD

NZD

PLN

CZK

HUF

RON

CHF

SEK

NOK

DKK

TRY

AED

SAR

ILS

ZAR

$0.00

Quantity

Discover similar items

Page 1 of 2

Robust testing capabilities

  • Broaden your test offerings with more than 120 different Beckman Coulter system reagent applications available
  • Conduct quick priority testing using the AU 480 clinical chemistry analyzer’s STAT wheel
  • Turn around critical samples fast with sample pre-dilution and auto-repeat of over-range results

Efficient reagent and consumable use

  • Minimize hands-on time and reduce daily reagent inventory tasks with ready-to-use liquid concentrated reagents
  • Experience lower total cost of ownership with low consumable requirements

Convenient operational features

  • Provide the right operator support when needed using context-sensitive help screens and instructional videos
  • Simplify operations with user-friendly graphical user interface design and touch-screen monitor
  • Reduce tech-to-tech variation and achieve consistent result quality with AU family of chemistry analyzers' standardized processes

    Analytical System

    Fully automated, random access chemistry system with STAT capability

    Analytical Principle

    Spectrophotometry and potentiometry

    Analytical Types

    Endpoint, rate, fixed point and indirect ISE

    Analytical Methods

    Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE

    Simultaneously Processed Analytes

    60 photometric tests + 3 ISE, 120 pre-programmed onboard tests

    Throughput

    400 photometric tests/hour; maximum of 800 with ISE

    ISE Throughput

    200 samples/hr. Maximum tests/hr: 600 if ISE only

    Sample Types

    Serum, plasma, urine, other

    Sample Feeder

    Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading

    Sample Tubes

    Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups

    STAT samples

    Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability.

    Sample Volume

    1–25 μl in 0.1 μl steps (1-25 μl for repeats)

    Reagent Capacity

    76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles.

    Reagent Storage

    Refrigerated (4–12°C)

    Reagent Volume

    R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment)

    Total Reaction Volume

    90–350 μl

    Reaction Cuvette

    Permanent glass cuvettes

    Reaction Time

    Up to 8 minutes, 37.5 seconds

    Reaction Incubation

    37°C, dry bath

    Photometry System

    Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible

    Wavelength

    13 different wavelengths between 340–800 nm

    Calibration

    Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode

    Quality Control

    Auto QC, cooled QC positions

    Rerun

    1. Auto rerun and manual rerun available
    2. Automatic sample increase, decrease or normal repeat
    3. Reflex testing

    Sample Pre-dilution

    Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100

    Test Requisition

    Individual and profile test requisition via online, mouse, function keys and touch-screen

    Sample Integrity

    Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection.

    Online

    RS232C. Full uni- and bidirectional communication possible.

    Software

    Windows 7®

    Data Storage

    Sample number 100,000 samples. Reaction Data 200,000 tests.

    Dimensions (W x H x D)

    ANL: 1450 mm x 1205 mm x 770 mm

    Power Supply

    100–240v; 60 Hz/ < 3.5 kVA